Version 1.14 – Technically Confirmed, now Clinically Feasible (Stage 3). November 18, 2016; updated June 15, 2023.
When referencing this document, please use the following format:
FDG-PET/CT Technical Committee. FDG-PET/CT as an Imaging Biomarker Measuring Response to Cancer Therapy, Quantitative Imaging Biomarkers Alliance, Version 1.14, Technically Confirmed Version. QIBA, November 18, 2016, updated June 15, 2023. Available from: RSNA.ORG/QIBA.
1. Executive Summary
This QIBA Profile documents specifications and requirements to provide comparability and consistency for quantitative FDG-PET across scanners in oncology. It can be applied to both clinical trial use and individual patient management, and organizes acquisition, reconstruction and post-processing, analysis and interpretation as steps in a pipeline that transforms data to information to knowledge.
The document was developed by the QIBA FDG-PET Biomarker Committee and shares content with the FDG-PET UPICT protocol, with additional material focused on the devices used to acquire and analyze FDG-PET data. Profile Part 3 is largely derived from the FDG-PET UPICT protocol, which uses a tiered hierarchy of compliance levels – Acceptable, Target, and Ideal. This Profile draws on the Acceptable components of the UPICT Protocol; later revisions are expected to draw on the Target and Ideal categories as the field advances.
The QIBA FDG-PET/CT Profile defines behavioral performance levels and quality-control specifications for whole-body FDG-PET/CT scans used in single- and multi-center clinical trials of oncologic therapies. While the emphasis is on clinical trials, the process is also intended to apply to clinical practice. The aim is to minimize intra- and inter-subject, intra- and inter-platform, and inter-institutional variability of quantitative scan data due to factors other than the intervention under investigation.
Note that specifications stated as ‘requirements’ in this document are only requirements to achieve the claim, not requirements on standard of care. Meeting the goals of this Profile is secondary to properly caring for the patient.
2. Clinical Context and Claims
FDG is a glucose analogue. Its use in oncology is based on the typically increased rate of glycolysis in tumors compared to normal tissue. FDG is transported into tumor cells via glucose transport proteins, usually up-regulated in tumor cells; once internalized it is phosphorylated to FDG-6-phosphate and becomes substantially metabolically trapped. FDG uptake is not specific for tumor cells – some normal tissues and processes with increased glucose turnover (e.g., infection and inflammation) also show elevated uptake.
This QIBA Profile specifically addresses the requirements for measurement of tumor FDG uptake with PET/CT as an imaging biomarker for evaluating therapeutic response. FDG-PET scans reliably reflect glucose metabolic activity of cancers, and this activity can be measured with high reproducibility over time. Longitudinal changes in tumor FDG accumulation during therapy often predict clinical outcomes earlier than changes in standard anatomic measurements.
Conformance to this Profile by all relevant staff and equipment supports the following claims:
Claim 1: Tumor glycolytic activity as reflected by the maximum standardized uptake value (SUVmax) is measurable from FDG-PET/CT with a within-subject coefficient of variation of 10–12%.
Claim 2: A measured increase in SUVmax of 39% or more, or a decrease of −28% or more, indicates that a true change has occurred with 95% confidence.
Considerations
- The Claim applies only to tumors considered evaluable with PET – in practice, tumors of a minimum size of 2 cm and a baseline SUVmax of 4 g/ml. More detail is given in Section 3.6.5.3.
- Details of the claim were derived from a review of the literature and are summarized in Appendix B.
- The asymmetric limits of repeatability arise because test-retest SUVmax differences do not follow a Normal distribution, but the differences of the logarithms of the test-retest values do. Converting the symmetric log-domain limits back to SUV units by exponentiation (RC = 100(exp(±1.96·SD(d)) −1)) yields necessarily asymmetric limits. Assuming a wCV of 12% for SUVmax gives repeatability coefficients of (−28%, +39%). For example, SUVmax values of 7.0 and 9.75 represent +39% as an increase or −28% as a decrease.
- The Claim is applicable for single-center studies using the same scanner. For multi-center studies, if imaging is performed using the same scanner and protocol for each patient at each time point, it is anticipated the Claim will be met.
- The Claim is based on SUVmax due to the evidence in the literature; however, SUV metrics derived from larger regions-of-interest (e.g., SUVpeak) are encouraged as they may provide improved repeatability.
- While informed by an extensive literature review, the Claim is currently a consensus Claim not yet substantiated by studies strictly conforming to these specifications, and should be re-assessed for technology changes such as PSF-based reconstruction or TOF imaging.
3. Profile Details
The assay method for computing and interpreting glycolytic metabolic activity using PET/CT may be viewed as a pipeline using either one or two or more scan sequences. The measure SUVx refers to one of several possible SUV measures, such as SUVmax, SUVmean or SUVpeak, with normalization by body weight or lean body mass.
The imaging steps are: (1) patients are prepared for scanning (e.g., 6 hr fasting), FDG is administered, and the patient waits quietly for bio-distribution and uptake of FDG (typically 60 min); (2) scan data from the PET and CT exams are acquired; (3) data correction terms are estimated and PET (and CT) images are reconstructed; (4) quantitative measurements are performed; (5) images are reviewed for qualitative interpretation. Steps 4 and 5 may occur in either order or at the same time.
The following table summarizes the major components and who performs them.
| Section | Title | Performed by |
|---|---|---|
| 3.1 | Subject Handling | Personnel (including Technologists and Schedulers) at an Image Acquisition Facility |
| 3.2 | Image Data Acquisition | Technologist, at an Image Acquisition Facility using an Acquisition Device |
| 3.3 | Image Data Reconstruction | Technologist, at an Image Acquisition Facility using Reconstruction Software |
| 3.4 | Image Analysis | Imaging Physician or Image Analyst using one or more Analysis Software tools |
| 3.5 | Image Interpretation | Imaging Physician before or after information obtained by Image Analysis using a pre-defined Response Assessment Criteria |
3.1. Subject Handling
This Profile refers primarily to ‘subjects’, keeping in mind that the recommendations apply to patients in general.
3.1.1 Subject Selection, Timing, and Blood Glucose Levels
The study protocol should include specific directions for managing subjects with abnormal fasting blood glucose, whether known to be diabetic or not. High levels of circulating blood glucose reduce FDG uptake, but there is a paucity of data to suggest a specific cutoff. Consideration should be given to excluding subjects with abnormal fasting blood glucose when quantitative FDG-PET/CT is the study’s primary endpoint. The study protocol should also define acceptable time intervals separating the FDG-PET/CT scan from the index intervention and other interventions, and an acceptable timing window around each imaging time point. Activities, tests and interventions that might increase the chance of false-positive or false-negative studies (e.g., biopsy-related inflammation, glucose tolerance tests, vigorous exercise) should be avoided prior to scanning.
3.1.2 Subject Preparation
The main purpose of subject preparation is to reduce tracer uptake in normal tissue (kidneys, bladder, skeletal muscle, myocardium, brown fat) while maintaining and optimizing uptake in target structures (tumor tissue). Key points include: subjects should not eat any food (oral or parenteral) for at least six hours prior to FDG administration; adequate hydration (glucose- and caffeine-free) is important; and strenuous or extreme exercise should be avoided for at least 6 hours (preferably 24 hours) before the exam. Upon arrival, confirmation of subject compliance with pre-procedure instructions and any potentially confounding events should be documented. Waiting and preparation rooms should be relaxing and warm (> 75° F / 22° C) during the uptake period, and the subject should remain recumbent or comfortably seated with activity and conversation kept to a minimum.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Height and Weight | Imaging Technologist | The Technologist shall measure and document subject height and weight and enter this information into the scanner during the PET/CT acquisition. Subject body weight shall be measured at the time of each PET/CT scan with standardized measurement devices and with the subject in an examination gown or light clothing. Subject height shall be measured and documented at the time of baseline FDG-PET scan with a standardized measurement device. Measurement of subject height is not required at each subsequent time point unless other subject-centric factors (e.g., growth in pediatric population or shrinkage in elderly population) are relevant in combination with a prolonged interval between imaging time points such that a change in height might be significant. If the subject cannot be moved from the bed, the date and source of information should be documented. The Technologist shall enter this information into a common data format mechanism used for recording all needed information (Appendix E). |
Diabetic Monitoring and Management (UPICT Section 4.2.2)
The subject’s blood glucose level should be measured [using a CLIA-approved, CLIA-cleared, or equivalent (outside US) glucose measurement device or laboratory] within the preceding 2 hours (ideally within 1 hour, especially in subjects with diabetes) of FDG administration and documented.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Blood glucose level measurement | Imaging Technologist or Lab Technologist | Within 2 hours preceding FDG administration, shall measure and document time of subject blood glucose collection. Glucose measurement should be performed using a CLIA-approved, CLIA-cleared, or equivalent (outside US) glucose measurement device. If the measurement is always performed at injection time or otherwise according to a protocol relative to injection time, this may be documented in the protocol. Deviations from this process shall be documented. |
| Blood glucose level documentation | Imaging Technologist or Lab Technologist | Shall enter the results of the blood glucose assay and the time of blood draw on a case report form or similar subject information sheet. Shall enter the results of the blood glucose assay into a common format mechanism used for recording all needed information (Appendix E). |
| Blood glucose level threshold | Imaging Technologist | Shall enforce the glucose thresholds for imaging as defined in the Protocol; if not, then reason for non-conformance shall be provided and documented on case report form or similar subject information sheet. Shall document any information on non-conformance with the protocol into a common format mechanism used for recording all needed information (Appendix E). |
3.1.3 Imaging-related Substance Preparation and Administration (UPICT Section 5)
FDG should be of high quality and purity, produced under Current Good Manufacturing Practice as specified by the FDA, EU, European Pharmacopeia, or other appropriate national regulatory agency. The 18F-FDG activity administered ranges between about 185–740 MBq (5–20 mCi), depending on the local imaging protocol. The exact activity and the time at which activity is calibrated should be recorded, along with any residual activity remaining in tubing, syringe, or administration system, so as to record the net amount of FDG injected into the subject.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Administered FDG Radiotracer Activity | Imaging Technologist | The Technologist shall: (1) Assay the pre-injection FDG activity (i.e., radioactivity) and time of measurement; (2) Inject the FDG as prescribed in the protocol, within the range defined in the protocol; (3) Record the time that FDG was injected into the subject; (4) Assay the residual activity in the syringe (and readily available tubing and components) after injection and record the time of measurement. These values shall be entered into the scanner during the PET/CT acquisition. For scanners that do not provide for entry of residual activity information, the net injected radioactivity should be manually calculated by decay-correcting all measurements to the time of injection and then subtracting the residual radioactivity from the pre-injection radioactivity. The net injected radioactivity is then entered into the scanner during the PET/CT acquisition. All data described herein on activity administration shall be documented. All data should be entered into the common data format mechanism (Appendix E). |
Radiotracer Administration Route (UPICT Section 5.4)
FDG should be administered intravenously through a large bore (24 gauge or larger) indwelling catheter placed anatomically remote to any site(s) of suspected pathology, preferably in an antecubital vein. If an infiltration or extraneous leakage is suspected, the event and expected quantity should be recorded and the infiltration site imaged. If the infiltration is greater than 5% of the administered activity and the quantitative result is a primary or secondary endpoint, the data point might be censored or the subject excluded.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| FDG Administration | Technologist | Technologist shall administer FDG intravenously through a large bore (24 gauge or larger) indwelling catheter placed anatomically remote to any sites of suspected pathology, preferably in an antecubital vein. Intravenous ports should not be used, unless no other venous access is available. In the case of manual administration, a three-way valve system should be attached to the intravenous cannula so as to allow at least a 10 cc normal (0.9% NaCl) saline flush following FDG injection. For automated injection devices alternate flushing mechanisms are allowed. |
| Suspected infiltration or extraneous leakage | Technologist and/or Physician or Physicist | Technologist shall document any suspected infiltration, leakage, or external contamination and consider scanning the injection site and/or contaminated materials. Record the event and expected amount of FDG into the common data format mechanism (Appendix E). |
3.1.3.2 CT Contrast Material Preparation and Administration
The presence of IV and/or oral contrast material improves lesion detection with CT and may improve anatomic localization, interpretation, and analysis. However, contrast material may affect the attenuation correction of the PET scan with consequent bias in measured SUVs. For each subject, the same approach should be followed for all imaging time points.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| CT Contrast Agent | Technologist | Technologist shall record the type and amount of CT Contrast Agent. (1) Was oral contrast used: Type [Positive, Negative], amount (volume in cc). (2) Was IV contrast used: Amount (volume in cc), time of injection. Record the event and expected amount of CT Contrast Agent into the common data format mechanism (Appendix E). |
3.2. Image Data Acquisition
This section summarizes the imaging protocols and procedures that shall be performed for an FDG-PET/CT exam. The motivation for controlling image acquisition tightly is that, over the course of a trial, hardware and software updates will occur; the intent is to ensure the instrument gives the same results over the duration of the trial. For consistency, clinical trial subjects should be imaged on the same device over the entire course of a study, and follow-up scans should be performed with identical acquisition parameters as the baseline, inclusive of all parameters required for both the CT and PET acquisitions.
Two scanning strategies are described. Strategy 1 uses the CT for attenuation correction and localization only (not a clinically diagnostic CT): CT scout, then CT for anatomic localization and attenuation correction, then PET emission acquisition. Strategy 2 applies where a clinically diagnostic CECT is also required, with option 2a (follow Strategy 1, then acquire an additional IV contrast-enhanced diagnostic CT) preferred since it avoids IV contrast impact on attenuation correction.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Scanning Strategy (Workflow) | Technologist | Technologist shall follow Profile-compliant workflow strategy, which will be compatible with Acquisition Device capability. The same workflow used at baseline shall be used at all subsequent time points. |
CT Exam Variables and Specifications
Positive contrast agents (IV or oral) may affect SUV quantitation by affecting the CT attenuation map. Each protocol should specify the desired approach; most importantly, for each subject the same approach should be followed for all imaging time points.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| CT Contrast agent | Technologist | CT contrast agents shall be given commensurate with the workflow strategy as selected from above. |
3.2.1.1 Timing of Image Data Acquisition
It is extremely important that (1) the time interval between FDG administration and the start of emission scan acquisition is consistent, and (2) when repeating a scan on the same subject, the same interval between injection and acquisition is used across time points. While the target tracer uptake time is 60 minutes, the acceptable window is from 55 to 75 minutes. When performing a follow-up scan, the same time interval should be applied with a target window of ±10 minutes, provided the scan does not begin prior to 55 minutes after injection.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Tracer Injection Time | Technologist | The time of FDG injection shall be entered into PET/CT scanner console during the acquisition. |
| Tracer Uptake Time | Technologist | The Technologist shall ensure that the tracer uptake time for the baseline scan is 60 minutes, with an acceptable range of 55 to 75 minutes. When repeating a scan on the same subject, especially in the context of therapy response assessment, the Technologist shall apply the same time interval ±10 minutes provided that the scan must not begin prior to 55 minutes after the injection of FDG. |
3.2.1.2 Subject Positioning (UPICT Section 7.2.1)
Consistent positioning avoids unnecessary variance in attenuation, gravity-induced shape and fluid distribution, and anatomical shape changes due to posture. Subjects should be positioned in the center of the field of view, preferably with arms positioned over head for whole-body imaging. Respiratory motion causes SUV errors by motion blurring and by errors in attenuation correction due to mismatches between the CT-based attenuation map and emission data. Shallow breathing shall be performed during CT AC acquisition, and it is of utmost importance that the patient not be in the end-inspiration phase during CT for PET/CT.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Subject Positioning | Technologist | The Technologist shall position the subject according to the UPICT specifications and/or specific protocol specifications consistently for all scans. |
| Positioning Non-conformance | Technologist | The Technologist shall document issues regarding subject non-conformance with positioning. The Technologist shall document issues regarding subject non-conformance with breathing and positioning using the common data format mechanism (Appendix E). |
| Respiratory motion minimization | Technologist | If the patient is observed to take a deep breath during the CT scan it should be documented and a repeat CT study should be considered. |
| Respiratory motion minimization | PET/CT Scanner | The PET/CT scanner shall provide methods to minimize the PET image errors introduced by respiratory motion. |
| Breathing and motion non-conformance | Technologist | The Technologist shall document issues regarding subject non-conformance with breathing and motion. The Technologist shall document issues regarding subject non-conformance with breathing and motion using the common data format mechanism (Appendix E). |
3.2.1.3 Scanning Coverage and Direction (UPICT Section 7.1.1)
For most oncology indications, anatomic coverage should include from the skull base (external auditory meatus) to the mid-thigh. Scanning caudocranial minimizes effects from increasing bladder activity during the scan. For a given subject, scanning direction on baseline scans must be duplicated at follow-up time points.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Scanning Direction | Technologist | The Technologist shall scan the subject caudocranial for whole body examination unless otherwise specified by the protocol. Each patient should be scanned consistently (in the same direction) over all time points. The scanning direction shall be entered into the PET/CT console during the acquisition and will be recorded by the scanner into the appropriate DICOM field. |
| Anatomic Coverage | Technologist | The Technologist shall perform the scan such that the anatomic coverage is acquired according to the protocol specifications and the same for all time points. |
3.2.1.4 Scanner Acquisition Mode Parameters
Acquisition mode parameters are those specified by the Technologist at the start of the actual PET/CT scan, including acquisition time per bed position, bed overlap, acquisition mode (2D or 3D), gating, and CT technique.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| PET acquisition mode | Study Sponsor | The key PET acquisition mode parameters (e.g., time per bed position, acquisition mode, with or without gating) shall be specified in a manner that is expected to produce acceptably low image noise regardless of the scanner make and model. The key acquisition mode parameters shall be specified according to pre-determined harmonization parameters. |
| PET acquisition mode | Technologist | The key PET acquisition mode parameters (e.g., time per bed position, acquisition mode, with or without gating) shall be set as specified by study protocol and used consistently for all patient scans. |
| Parameter | Entity/Actor | Specification |
|---|---|---|
| CT acquisition mode | Study Sponsor | The key CT acquisition mode parameters (kVp, mAs, pitch, and collimation) shall be specified in a manner that is expected to produce comparable results regardless of the scanner make and model and with appropriate radiation doses consistent for the role of the CT scan: diagnostic CT scan, anatomical localization, or accurate corrections for attenuation and scatter. CT dose reduction techniques may be used if demonstrated to retain quantitative accuracy. |
| CT acquisition mode | Technologist | The key CT acquisition mode parameters (kVp, mAs, pitch, and collimation) shall be set as specified by study protocol and used consistently for all subject scans. |
Regarding CT radiation exposure, the lowest radiation dose necessary to achieve the diagnostic objective should be used. Higher kVp CT acquisitions are recommended in general, as lower kVp acquisitions can be more biased for estimating PET tracer uptake in bone. Protocols should allow high enough mAs to avoid bias from extremely low x-ray flux.
3.3. Imaging Data Reconstruction and Post-Processing
Reconstructed image data is the PET image exactly as produced by the reconstruction process on the PET/CT scanner – a stack of DICOM slices constituting a PET image volume with no processing other than that occurring during reconstruction. PET reconstruction parameters include the choice of algorithm, number of iterations and subsets, type and amount of smoothing, field of view, and voxel size. Standardization of reconstruction settings is necessary to obtain comparable resolution and SUV recoveries across subjects and across sites.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| PET image reconstruction | Study Sponsor | The key PET reconstruction parameters (algorithm, iterations, smoothing, field of view, voxel size) shall be specified. The image voxel size should be <5 mm (strongly prefer 3–4 mm), with <3 mm voxels for head and neck imaging. The key PET image reconstruction parameters shall be specified according to pre-determined harmonization parameters. |
| PET image reconstruction | Technologist | The key PET reconstruction parameters (algorithm, iterations, smoothing, field of view, voxel size) shall be followed and set as specified. |
| Correction factors | Technologist | All quantitative corrections shall be applied during the image reconstruction process. These include attenuation, scatter, randoms, dead-time, and efficiency normalizations. |
| Calibration factors | Scanner | All necessary calibration factors needed to output PET images in units of Bq/ml shall be automatically applied during the image reconstruction process. |
3.3.2 Image Data Post-processing (UPICT Section 8)
Processed image data are images transformed in some manner – including smoothing, zoom, rotation/translation, resampling, interpolation, slice averaging, or MIP. Where image registration or interpolation is required, the ROI analysis should be performed on the original PET image set using appropriately modified ROIs, to preserve the numerical accuracy of the true PET image values. Standard whole-body FDG-PET oncology studies typically include all necessary corrections within reconstruction and require no additional processing other than de-identification.
3.3.3 Imaging Data Storage and Transfer
The term ‘raw data’ is ambiguous: it can refer to scanner raw data (sinograms or list-mode) or image raw data. For archiving at the local site or imaging core lab, the most important data are the original images (image raw data).
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Data archiving | Technologist | The originally reconstructed PET images (image raw data), with and without attenuation correction, and CT images shall always be archived at the local site. If processed PET images are required, they shall be archived as separate secondary datasets. If scanner raw data need to be archived for future reprocessing, this should be defined prospectively in the Protocol. |
3.4. Image Analysis (UPICT Section 9)
The Image Analyst, through interaction with the Workstation Analysis tools, shall be able to perform specified measurements. The output images of Reconstruction, but not Post-processing, are considered the input for Image Analysis. Each tissue/organ to be investigated quantitatively is characterized by defining a region-of-interest (ROI) and calculating a parameter such as the maximum SUV within the ROI.
The specific trial protocol shall prospectively define the SUV parameter required for each lesion or normal tissue used for the imaging endpoint. SUV measures (and the analysis tools used to obtain them, including software version) shall be specified for each protocol and used consistently across all subjects and all sequential lesion measurements. SUVs are intended as a measure of relative uptake; under the common body-weight normalization the units are g/ml. It should be reported which SUV measure is used (statistic and type of normalization), and if a reference tissue (e.g., liver) SUV is measured, it should be reported along with lesion SUV data.
3.5. Image Interpretation and Reporting (UPICT Section 10)
No QIBA Profile specification can be provided for image interpretation at this time; image interpretation is considered beyond the scope of this document. Further interpretation of the quantitative results (e.g., PERCIST) and/or normalizing SUV to reference-tissue values (e.g., liver or blood pool) can be specified as part of a specific trial protocol. Typically the trial protocol will state how quantitative response is measured – for example, based on the hottest lesion, or the change in the sum of SUVs – and this should be specified a priori by the trial itself.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Image Reporting | Imaging Facility | Imaging reports shall be populated from DICOM header information using structured reporting. |
3.6. Quality Control
This section addresses multiple aspects of quality control in FDG-PET/CT studies, including selecting and qualifying a PET/CT imaging facility, imaging personnel, PET/CT scanners and ancillary equipment, phantom imaging, and post-image-acquisition quality assessment.
3.6.1 Imaging Facility
Quality processes that ensure reliable scanner performance and consistent image acquisition methodology must be in place prior to subject imaging and followed for the duration of the trial. A facility “imaging capability assessment” is a prerequisite to facility selection. Imaging facility Accreditation and/or Qualification are considered necessary but not sufficient for conformance with this Profile; for conformance, all normative requirements must be met.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Accreditation / Qualification | Imaging Site & Image Acquisition Device | Shall maintain and document Accredited status for clinical practice (ACR, IAC, TJC, etc.) or Qualified status for clinical trials (e.g. ACRIN, SNM-CTN, CALGB, CROs, etc.). |
3.6.2 Imaging Facility Personnel
For each personnel category, training, credentialing, continuing education and peer review standards should be defined.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Personnel Roster | Imaging Facility Coordinator | Each site shall, at the time of trial activation and prior to subject accrual, have the support of certified technologists, physicists, and physicians (as defined below), experienced in the use of FDG-PET/CT in the conduct of oncological clinical trials. |
| Technologist | Imaging Facility Coordinator | Technologist certification shall be equivalent to the recommendations published by the representatives from the Society of Nuclear Medicine Technologists Section (SNMTS) and the American Society of Radiologic Technologists (ASRT) and should also meet all local, regional, and national regulatory requirements for the administration of ionizing radiation to patients. |
| Medical Physicist | Imaging Facility Coordinator | Medical physicists shall be certified in Medical Nuclear Physics or Radiological Physics by the American Board of Radiology (ABR); in Nuclear Medicine Physics by the American Board of Science in Nuclear Medicine (ABSNM); in Nuclear Medicine Physics by the Canadian College of Physicists in Medicine; or equivalent certification in other countries; or have 3 years of PET experience. Regardless of certification, the physicist should have specific experience in PET and its quantitative use. |
| Physician | Imaging Facility Coordinator | Physicians overseeing and interpreting PET/CT scans shall be qualified by the ABR (Diagnostic and/or Nuclear Radiology) or American Board of Nuclear Medicine (ABNM) or equivalent within the United States or an equivalent entity appropriate for the geographic location in which the imaging study(ies) will be performed and/or interpreted. |
3.6.3 FDG-PET/CT Acquisition Scanner
FDG-PET/CT studies as described in this Profile require a dedicated PET/CT scanner, identified by manufacturer, name and model, with hardware specifications and software version documented at trial initiation and at all updates or upgrades.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Physical Inspection | Technologist | Shall, on a daily basis, check gantry covers in tunnel and subject handling system. |
| QA/QC Checks | Technologist | All QA/QC procedures recommended by the manufacturer shall be performed at the frequency recommended by the manufacturer and documented. A table of QA/QC procedures for a subset of specific PET/CT scanners from each vendor is included in Appendix G.2. Daily QA procedures shall be performed prior to any subject scan. |
3.6.3.1 Ancillary Equipment – Dose Calibrator
The Constancy test ensures reproducibility of an activity measurement over a long period. The Accuracy test ensures that activity values are correct and traceable to national or international standards. The Linearity test confirms that the same calibration setting can be applied to obtain the correct activity readout over the range of use.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Constancy | Technologist | Shall be evaluated daily (or after any dose calibrator event) using a NIST-traceable (or equivalent) simulated F-18, Cs-137, or Co-57 dose calibrator standard and confirmed that net measured activity on the F-18 setting differs by no greater than ±2.5% from day to day. |
| Accuracy | Technologist | Shall be evaluated monthly (or after any dose calibrator event) with a NIST-traceable (or equivalent) Cs-137, Co-57, or simulated F-18 dose calibrator standard. Shall confirm that net measured activities differ no greater than ±2.5% from expected value. The scanner calibration shall be tested using a NIST-traceable (or equivalent) simulated F-18 source object, e.g., a uniform cylinder, large enough to avoid partial volume effects or other resolution losses. |
| Linearity | Technologist or Radiation safety officer or Qualified Medical Physicist | Shall be evaluated at least annually (or after any dose calibrator event) and should be within ±2.5% of the true value over an operating range of 37–1110 MBq (1 to 30 mCi). |
| PET Radiation Dose | Dose Calibrator | Shall record the radiation dose from the administered activity and accompanying information in a DICOM Radiopharmaceutical Administration Radiation Dose Structured Report. |
3.6.3.1.2 Scales and Stadiometers
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Scales and stadiometers | Approved personnel | Shall be evaluated annually or after any repair by qualified personnel. Shall be confirmed that error is less than ±2.5% from expected values using NIST-traceable or equivalent standards. |
3.6.3.1.3 Blood Glucose Level Measurement Device
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Blood glucose level measurement device | Approved personnel | Shall have QA/QC testing and calibration performed using a CLIA-approved, CLIA-cleared, or equivalent (outside US) procedure. |
3.6.3.1.4 Clocks and Timing Devices
PET/CT scanner computer and all clocks used to record activity/injection measurements should be synchronized to standard time reference within ±1 minute. Correct synchronization could be achieved using the IHE Consistent Time Integration Profile, which requires the Network Time Protocol (NTP).
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Scanner and site clocks | Approved personnel | PET/CT scanner computer and all clocks in an imaging facility used to record activity/injection measurements shall be synchronized to standard time reference within ±1 minute. Synchronization of all clocks used in the conduct of the FDG-PET/CT study shall be checked weekly and after power outages or civil changes for Daylight Saving (NA) or Summer Time (Eur). |
| Scanner and site clocks | Specific Device | Provide time synchronization as per the IHE Consistent Time Integration Profile. |
| Dose calibrator clock | Dose Calibrator | Electronic record of output from a dose calibrator shall be synchronized with other time keeping devices. |
3.6.4 Phantom Imaging
To qualify the PET/CT scanner for clinical practice or a clinical trial, a phantom imaging procedure is required. Phantom scans and performance evaluation should be performed prior to the start of a trial and repeated during the course of the trial as specified by the individual protocol. Image noise is measured using an anthropomorphic phantom with a uniform area, assessed by the coefficient of variation (COV = SD/Mean × 100). The phantom should be filled to an activity concentration of approximately 3.7–7.4 kBq/ml (0.1–0.2 µCi/ml), and the COV of voxel values within a region no smaller than 3 cm to a side should be below 15%.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Phantom tests: Frequency | Imaging Site | Shall perform and document results of the following tests no less than quarterly. |
| Phantom tests: cross calibration with dose calibrator | Imaging Site | Shall perform quarterly and after scanner upgrades, maintenance or repairs, new setups and modifications to the dose calibrator. |
| Phantom tests: SUV measurements | Imaging Site | Using ACR or uniform cylinder phantom or equivalent shall obtain an SUV for a large central ROI of 1.0 with an acceptable range of 0.9 to 1.1. |
| Phantom tests: axial uniformity measurements | Imaging Site | Using uniform cylinder phantom or equivalent shall obtain a slice-to-slice variability of less than 10% for the slices within the central 80% of the axial FOV. |
| Phantom tests: resolution measurements | Imaging Site | Shall perform and successfully obtain phantom imaging results for cold and hot object imaging as described. For cold object imaging, the test phantom will be the ACR PET phantom or the Deluxe Jaszczak phantom (or equivalent) with six sets of acrylic rods arranged in a pie-shaped pattern with the following diameters: 4.8, 6.4, 7.9, 9.5, 11.1, and 12.7 mm. The 9.5, 11.1, and 12.7 mm diameter rods must be visible; if necessary up to 1 cm slice averaging can be used. For hot object resolution, the fillable 12 mm diameter ‘hot’ cylinder for the ACR phantom must be visible; for CTN phantom, all hot objects greater than 10 mm must be visible and for NEMA phantom, 13 mm sphere must be visible. Also see Section 3.6.4.2. |
| Phantom tests: noise measurements | Imaging Site | The phantom shall be filled with an FDG concentration of activity concentration in the uniform area of approximately 0.1 to 0.2 µCi/ml and scanned using the intended acquisition protocol. Using a rectangular or spherical region as close as possible to, but no smaller than, 3 cm to a side, the COV of the voxel values within the region should be below 15%, for the slices within the central 80% of the axial FOV. |
3.6.4.1 Uniformity and Calibration
Verification of scanner normalization with a uniform phantom is a minimum requirement for all scanners used in clinical trials, including those with only qualitative endpoints. For trials with quantitative measurements, this assessment should include a comparison against a dose calibrator (within 10%) to ensure quantitative accuracy.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Uniformity QC | Technologist or Physicist | At least quarterly and following software upgrades, shall assess transverse and axial uniformity across the central 80% of image planes by imaging a uniform cylinder phantom. (1) The standard deviation of a large central 2D ROI shall be compared with similar previous scans to check for measurable differences. (2) The mean values of a large central 2D ROI for these image slices shall be compared with similar previous scans to check for measurable differences. |
| Cross Calibration | Technologist or Physicist | At least quarterly and following software upgrades or changes to the dose calibrator, shall perform checks to monitor and identify discrepancies between the PET scanner and dose calibrator. |
3.6.4.2 Resolution (UPICT Section 12.1.1.11)
Assessment of adequate resolution should include both a qualitative evaluation (using clinical images) and a quantitative assessment (using phantom-defined criteria).
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Resolution | Nuclear Medicine Physician | Shall perform, on at least an annual basis, and document a qualitative resolution QC test by using the routine image processing protocols and demonstrating resolution of normal gross anatomic features within clinical images of the brain, heart and abdomen. |
| Resolution | Medical Physicist | Shall perform (on at least an annual basis) and document performance of a quantitative assessment (using a phantom with differing size defined targets such as the ACR or NEMA IQ phantoms processed with routine image reconstruction protocols) for lesion resolution. |
3.6.4.3 Noise (UPICT Section 12.1.1.12)
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Noise | Medical Physicist | Shall perform qualitative or quantitative assessment of image noise in phantom images to be of consistent and acceptable quality. |
3.6.4.4 Phantom Imaging Data Analysis
To provide a method for testing and validating quantitative accuracy of SUV measurements produced by display and analysis methods, the QIBA FDG-PET/CT Biomarker Committee developed an FDG-PET/CT digital reference object (DRO) – a synthetic test object comprised of stacked DICOM images representing an FDG-PET image volume and an aligned CT image volume, based on the NEMA/MITA NU-2 Image Quality phantom. Further details are given in Appendix F.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Image Analysis Software | Imaging site | All software used for image analysis shall be verified by the procedures in 4.4.3. |
3.6.5 Quality Control of FDG-PET/CT Studies
The integrity of DICOM image headers should be reviewed and confirmed for DICOM standard conformance, regulatory conformance, protocol conformance, and sufficiency for the intended analysis. CT images should be reviewed for image quality and potential artifacts (beam hardening, metal objects, motion); PET images should be compared to CT for proper registration and potential attenuation-correction artifacts.
Determination of Evaluable Tumor Lesions. The lesion to be measured should be free of artifacts. A high baseline FDG uptake is necessary to reliably measure change: a minimum FDG-avidity is required and should be specified in the protocol, either via a subject-specific threshold (as in PERCIST) or a general cutoff. For a general cutoff, an SUVmax of 4 is suggested for all target lesions, though a lower minimum may be acceptable in some settings (e.g., lung or breast). The measurement for mean liver SUV is made using a 3-cm diameter spherical ROI placed in the right lobe at the level of the main portal vein, equidistant between the porta hepatis and lateral liver margin. If the liver is not evaluable, the alternate comparator is 2.0 × mean SUV of blood pool in a 3D ROI (a 1 cm diameter cylinder in the descending thoracic aorta extending over 2 cm). Tumors should typically be over 2 cm in diameter for target lesion inclusion at baseline; smaller lesions with high enough FDG uptake may still be evaluable.
4. Conformance
Qualified: The imaging site is formally approved by an appropriate body (i.e., ACRIN, CQIE, SNMMI-CTN, EANM-EARL, NCRI, an imaging laboratory or CRO) for a specific clinical research study. Accredited: Approval by an independent body or group for broad clinical usage (requires ongoing QA/QC) e.g., ACR, IAC, TJC. Compliant: The imaging site and equipment meet all the requirements described herein, which are necessary to meet the QIBA Profile claim.
The requirements are intended to establish a baseline level of capabilities. Providing higher levels of performance or advanced capabilities is both allowed and encouraged. Institutions meeting the stated criteria are considered to be QIBA Compliant.
4.1 Image Acquisition Site
A QA/QC program for PET/CT scanners and ancillary devices must be in place to achieve the goals of the clinical trial. This program shall include elements to verify that imaging facilities are performing imaging studies correctly and that a facility’s PET/CT scanners are performing within specified calibration values.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| PET/CT Scanner | Acquisition Facility | This Profile shall only address full ring PET/CT scanners. |
| CT Scanner Calibration | Technologist | Shall perform daily water equivalent phantom analysis; ensure that output is acceptable and manually enter on form/electronic database. |
| PET Scanner Calibration | Technologist | Shall perform daily/weekly/monthly scanner QA as recommended by manufacturer; ensure that output values are acceptable and manually enter on form/electronic database. |
| PET Scanner Calibration Constancy Check | Technologist | Shall perform constancy phantom (e.g., Ge-68 cylinder) scan (preferably NIST traceable or equivalent to gather information regarding uniformity as well) at least weekly and after each calibration. |
| Dose calibrator | Imaging site or Manufacturer | Calibrated to F-18 using NIST traceable source or equivalent either by site or calibrator manufacturer. |
4.2. PET/CT Acquisition Device
Distinct from the performance specifications and frequency of testing in Section 4.1 (which apply to quality control of the Acquisition Device at the imaging facility), this Section defines performance specifications of the Acquisition Device to be met upon leaving the manufacturing facility. The PET scan acquisition start time should be used for the decay reference time and the integral model should be used for decay correction. The scanner should perform all decay corrections (not the operator), and image data are to be given in units Bq/ml. All needed information for fully corrected administered activity (residual activity, injection time, calibration time) is required.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| CT calibration tracking | Acquisition Device | Daily water equivalent phantom values shall be tracked in the DICOM header. |
| PET calibration factor | Acquisition Device | The current SUV calibration factor shall be included in the DICOM header. |
| PET QA status | Acquisition Device | Date/time and status of system-wide QA checks should be captured separately. |
| Dose calibrator calibration | Acquisition Device | Calibration factor for an F-18 NIST-traceable (or equivalent) source with identifying information shall be tracked in the DICOM header with Date/Time. |
| PET Scanner calibration | Acquisition Device | Shall be able to be calibrated according to the following specifications: Using a uniform cylinder containing F-18 in water (ideally the same used for dose calibrator cross-calibration). Slice-to-slice variability shall be no more than ±5% (not including end slices, as per ACR PET Core Lab). In-plane uniformity for the above phantom shall be less than 5%. Immediately after calibration, using the same filled phantom and scanned sufficiently long to minimize statistical noise, and an ACR-type ROI analysis, the average measured SUV shall be in the range of 0.98 to 1.02. (Note this is not the performance expected during clinical imaging operation as discussed in the preamble to this Section.) This technique removes the variability due to radionuclide measurement. |
| Weight | Acquisition Device | Shall be able to record patient weight in lbs or kg as supplied from the modality worklist or operator entry into scanner interface. Shall be stored in Patient Weight field (0010,1030) in the DICOM image header, as per DICOM standard. Patient weight shall be specifiable with 4 significant digits. Patient weight shall be transferrable directly from measurement device into scanner by electronic, HIS/RIS, or other means, bypassing all operator entry, but still permitting operator correction. |
| Height | Acquisition Device | Shall be able to record patient height in feet/inches or cm/m as supplied from the modality worklist or operator entry into scanner interface. Shall be stored in Patient Size field (0010,1020) in the DICOM image header, as per DICOM standard. Patient height shall be specifiable with 3 significant digits. Patient height shall be transferrable directly from measurement device into scanner by electronic, HIS/RIS, or other means, bypassing all operator entry, but still permitting operator correction. |
| Blood glucose level | Acquisition Device | Shall be able to record patient blood glucose level, in units of mg/dl or mMol/l, and time of measurement, as supplied by operator entry into the scanner interface. Shall be recorded in the DICOM image header in the Acquisition Context Sequence using DICOM PS 3.16 TID 3471 PET Covariates Acquisition Context. Patient blood glucose level shall be transferrable directly from measurement device into the scanner using the Modality Worklist NM/PET Protocol Context TID 15101, bypassing all operator entry, but displaying it to the operator and still permitting operator correction. |
| Administered Radionuclide | Acquisition Device | Shall be able to enter the radionuclide type (i.e., F-18) by operator entry into the scanner interface and through predefined protocol. Shall be recorded in Radionuclide Code Sequence (0054,0300) in the DICOM image header [e.g., (C-111A1, SRT, “18Fluorine”)]. Shall be able to accept the radionuclide type directly from the measurement device (dose calibrator) or management system, using the Sup 159 Radiopharmaceutical Administration Radiation Dose Report bypassing all operator entry, but still permitting operator correction. Shall be able to accept the radionuclide type from the DICOM Modality Worklist either from the NM/PET Protocol Context, if present, or by deriving it from the Requested Procedure Code via a locally configurable table of values. |
| Administered Radiotracer | Acquisition Device | Shall be able to record the radiotracer (i.e., FDG), as supplied by operator entry into the scanner interface. Shall be recorded in Radionuclide Code Sequence field (0054,0300) in the DICOM image header, e.g., (C-B1031, SRT, “Fluorodeoxyglucose F18”). |
| Administered Radiotracer radioactivity | Acquisition Device | Shall be able to enter the administered radioactivity, in both MBq and mCi, as supplied by operator entry into the scanner interface. Shall be recorded in Radionuclide Total Dose field (0018,1074) in the DICOM image header in Bq. Shall be able to record with separate entry fields on scanner interface: (1) the pre-injection FDG radioactivity; (2) time of measurement of pre-injection FDG radioactivity; (3) the residual activity after injection; (4) time of measurement of the residual radioactivity after injection. Shall automatically calculate the administered radioactivity and store in the Radionuclide Total Dose field (0018,1074). Alternatively, shall be able to receive this information as per DICOM Supplement 159. Patient Administered Radiotracer radioactivity information shall be transferred directly from measurement device into scanner by electronic, HIS/RIS, or other means, bypassing all operator entry, but still permitting operator correction. |
| Administered Radiotracer Time | Acquisition Device | Shall be able to record the time of the start of activity injection as supplied by operator entry into the scanner interface. Shall be recorded in Radiopharmaceutical Start Date Time field (0018,1078) (preferred) or Radiopharmaceutical Start Time field (0018,1072). Shall be able to record the time of the stop of activity injection. Shall be recorded in Radiopharmaceutical Stop Date Time field (0018,1079). |
| Decay Correction Methodology | Acquisition Device | Encoded voxel values with Rescale Slope field (0028,1053) applied shall be decay corrected by the scanner software (not the operator) to a single reference time (regardless of bed position), which is the start time of the first acquisition, which shall be encoded in the Series Time field (0008,0031) for original images. Corrected Image field (0028,0051) shall include the value “DECY” and Decay Correction field (0054,1102) shall be “START”, which means that the images are decay corrected to the earliest Acquisition Time (0008,0032). |
| Scanning Workflow | Acquisition Device | Shall be able to support Profile Protocol (Section 3) PET and CT order(s) of acquisition. Shall be able to pre-define and save (by imaging site) a Profile acquisition Protocol for patient acquisition. Shall be able to interpret previously-reconstructed patient images to regenerate acquisition protocol. Shall be configurable to store (or receive) acquisition parameters as pre-defined protocols (in a proprietary or standard format), to allow re-use of such stored protocols to meet multi-center specifications and to achieve repeatable performance across time points for the same subject. |
| CT Acquisition Parameters | Acquisition Device | Shall record all key acquisition parameters (technique) in the CT image header, using standard DICOM fields. |
| CT based attenuation correction | Acquisition Device | Shall record information in PET DICOM image header regarding which CT images were used for corrections (attenuation, scatter, etc.). |
| PET-CT Alignment | Acquisition Device | Shall be able to align PET and CT images within ±2 mm in any direction. Shall be able to align PET and CT images within ±2 mm in any direction with a load up to 150 kg over the co-scan length. |
| Activity Concentration in the Reconstructed Images | Acquisition Device | Shall be able to store and record (rescaled) image data in units of Bq/ml and use a value of BQML for Units field (0054,1001). |
| Tracer Uptake Time | Acquisition Device | Shall be derivable from the difference between the Radiopharmaceutical Date Time field (0018,1078) (preferred) or Radiopharmaceutical Start Time field (0018,1072) and the Series Time field (0008,0031) or earliest Acquisition Time field (0008,0032) in the series (i.e., the start of acquisition at the first bed position), which should be reported as series time field (0008,0031). |
| PET Voxel size | Acquisition Device | See Section 4.3 (PET Voxel size) under the Reconstruction Software specification requirements. |
| CT Voxel size | Acquisition Device | Shall be no greater than the reconstructed PET voxel size. Voxels shall be square, although are not required to be isotropic in the Z (head-foot) axis. Not required to be the same as the reconstructed PET voxel size. |
| Subject Positioning | Acquisition Device | Shall be able to record the subject position in the Patient Orientation Code Sequence field (0054,0410) (whether prone or supine) and Patient Gantry Relationship Code Sequence field (0054,0414) (whether head or feet first). |
| Scanning Direction | Acquisition Device | Shall be able to record the scanning direction (craniocaudal vs. caudocranial) into an appropriate DICOM field. |
| Documentation of Exam Specification | Acquisition Device | Shall be able to define the extent of anatomic coverage based on distance from defined landmark site (e.g., vertex, EAM) — both the landmark location (anatomically) and the distance scanned from landmark would require DICOM tags. Shall be able to be reportable for future scanning sessions. The Acquisition Device shall record the z-axis FOV which represents the actual distance of scan anatomic coverage (cm). |
| Differential Acquisition Time | Acquisition Device | Shall be able to acquire and record non-uniform scan times dependent upon areas of clinical concern. Recording can be done through the use of Actual Frame Duration (0018,1242) and Frame Reference Time (0054,1300). |
| DICOM Conformance | Acquisition Device | All image data and scan parameters shall be transferable using appropriate DICOM fields according to the DICOM conformance statement for the PET/CT scanner. |
| DICOM Data transfer and storage format | PET/CT Scanner or Display Workstation | PET images shall be encoded in the DICOM PET or Enhanced PET Image Storage SOP Class, using activity-concentration units (Bq/ml) with additional parameters stored in public DICOM fields to enable calculation of SUVs. PET images shall be transferred and stored without any form of lossy compression. |
| DICOM Editing | Acquisition Device | Shall be able to edit all fields relevant for SUV calculation and blood glucose before image distribution from scanner. Shall provide appropriate warnings if overriding of the current values is initiated. |
4.3. Reconstruction Software
Reconstruction Software shall propagate the information collected at the prior Subject Handling and Image Acquisition stages and extend it with those items noted in the Reconstruction section. Data can be reconstructed including all corrections needed for quantification as well as without scatter and attenuation correction. If the system is capable of providing resolution recovery and/or time of flight, the decision to turn these capabilities on or off should be made prospectively, as dictated by the specific protocol, and should be consistent for a given subject across multiple time points.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Metadata | Reconstruction Software | Shall be able to accurately propagate the information collected at the prior stages and extend it with those items noted in the Reconstruction section. |
| Data Corrections | Reconstruction Software | PET emission data must be able to be corrected for geometrical response and detector efficiency, system dead time, random coincidences, scatter and attenuation. |
| Reconstruction Methodology | Reconstruction Software | Shall be able to provide images without resolution recovery. Shall be able to indicate, for both TOF and Resolution recovery, if either are being used for purposes of image reconstruction. |
| Reconstruction Methodology / Output | Reconstruction Software | Shall be able to perform reconstructions with and without attenuation correction. |
| Data Reconstruction 2D/3D Compatibility | Reconstruction Software | Shall be able to perform reconstruction of data acquired in 3D mode using fully 3D image reconstruction algorithms. Shall be able to perform reconstruction of data acquired in 2D mode using 2D image reconstruction algorithms. |
| Quantitative calibration | Reconstruction Software | Shall apply appropriate quantitative calibration factors such that all images have units of activity concentration, e.g., kBq/mL. |
| Multi-bed data | Reconstruction Software | Shall combine data from multiple over-lapping bed positions (including appropriate decay corrections) so as to produce a single three dimensional image volume. |
| Voxel size | Reconstruction Software | Shall allow the user to define the image voxel size by adjusting the matrix dimensions and/or diameter of the reconstruction field-of-view. Shall be able to reconstruct PET voxels with a size 4 mm or less in all three dimensions (as recorded in Voxel Spacing field (0028,0030) and computed from the reconstruction interval between Image Position (Patient) (0020,0032) values of successive slices). Pixels shall be square, although voxels are not required to be isotropic in the z (head-foot) axis. |
| Reconstruction parameters | Reconstruction Software | Shall allow the user to control image noise and spatial resolution by adjusting reconstruction parameters, e.g., number of iterations, post-reconstruction filters. Shall be able to record reconstruction parameters used in image DICOM header using the Enhanced PET IOD, developed by DICOM working group. |
| Reconstruction protocols | Reconstruction Software | Shall allow a set of reconstruction parameters to be saved and automatically applied (without manual intervention) to future studies as needed. |
4.4. Image Analysis Workstation
The image analysis workstation shall be able to receive and propagate the data output (imaging and metadata) collected from the prior activities. The analysis workstation and software may be coupled to the PET/CT scanner system or provided by a 3rd-party vendor.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Metadata | Image Analysis Workstation | Shall be able to accurately propagate the information collected at the prior stages and extend it with those items noted in the Image Analysis Workstation section. Shall be able to display all information that affects SUVs either directly in calculation (e.g., patient weight, injected activity) or indirectly (uptake time, plasma glucose concentration). |
| Tracer Uptake Time: Display | Image Analysis Workstation | Shall be capable of displaying or including a link to display the number of minutes between injection and initiation of imaging (as per derivation guidelines described in Section 4.2). |
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Reference time for decay correction | Image Analysis Workstation | Shall use either the Acquisition Time field (0008,0032) or Radiopharmaceutical Start Time (0018,1072), if necessary. If a series (derived or not) is based on Acquisition Time decay correction, the earliest Acquisition Time (0008,0032) shall be used as the reference time for decay correction. |
4.4.1 Region of Interest (ROI) Definition
The scanner-display-analysis system shall provide a tool for the user to define both 2D and 3D regions of interest. While the ROI can be drawn on processed images, the SUV calculation should be performed from unprocessed (raw) image data. The specifications support calculation of (1) average value within an ROI (SUVmean), (2) maximum value within an ROI (SUVmax), (3) average value within a fixed-size ROI (SUVpeak), and (4) average value within a fixed-size ROI with location automatically selected to maximize the mean value. For SUVpeak measures, the use of partial voxel values to secure a 1.2 cm diameter sphere (or 1 cc volume) ROI is appropriate and desirable.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Voxel Inclusion Policy | Analysis Tool | Shall describe voxel inclusion methodology and weighting policy including placement criteria and total volume. Use a method equivalent to weighting for partial voxels; fully included voxels use weight of 1.0. Weighting should be proportionate to volumes of voxels that are partly included. |
| ROI Specifications | Analysis Tool | Shall describe capabilities and limits of ROI specification and placement. Dimensions and center location of ROI (box, ellipse, or ellipsoid) shall be specifiable to ±1 mm. For SUVpeak measures, the location within a target search region that yields the highest mean value of a 1 cc region shall be found automatically and reproducibly. |
| ROI Definition Tools | Analysis Tool | Shall provide a tool and user strategy to allow the placement of an ROI to determine the average value within the ROI. Shall provide a tool and user strategy to allow the placement of an ROI to determine the value and location of the voxel with the maximum value within an ROI. Shall provide a tool and user strategy to allow the placement of a 1 cm diameter ROI (either 2D or 3D) to determine the average value within the ROI. Shall provide a tool and user strategy to allow automatic placement of a 1 cm diameter ROI (either 2D or 3D) such that the average value within the ROI is maximized. |
| Edge/Volume Detection | Analysis Tool | Shall provide threshold methods for defining an ROI based on image values. Shall clearly specify which threshold method is used and relevant parameter values. Three ROI definition methods shall be provided: Fixed value, % of maximum voxel, or edge detection/segmentation methods. |
| ROI saving/retrieve | Analysis/Archival | Shall have the capability to label, save, recall and edit ROIs. Shall have the capability to track tumor information across longitudinal scans. In addition to lesion (and normal reference region) identification, this may include cross time point mapping of lesions tracked on the basis of consistent anatomic and/or functional activity. Other lesion characteristics, such as lesion name (with consistent anatomic labeling), lesion location, ROI/VOI size, corresponding anatomic (CT) image or slice number, SUV metric(s) and assessment of tumor heterogeneity may also be tracked and captured using standard DICOM objects. |
| ROI Display Statistics | Analysis Tool | Shall have the capability to output to the screen display the selected statistics of the ROI. These include, but are not limited to: area, volume, mean, maximum, minimum, standard deviation. Units can be selectable as activity concentration [Bq/ml] or SUV [g/ml] (See Section 3.4.3). Shall have the capability to display results with at least two decimal places. Shall output ROI Output Statistics to Structured Data Reporting DICOM files. Shall calculate results directly from the originally reconstructed voxels (not from interpolated and/or zoomed images). |
4.4.2 Calculation of Standardized Uptake Value (SUV)
The ROI definition and analysis software is responsible for SUV calculation, e.g., with decay correction to the appropriate reference time. The manufacturer should implement both versions of SUV normalizations (body weight or lean body mass). Recommended vendor-neutral pseudo-codes for SUV calculation are given in Appendix G.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| SUV Calculation | Analysis Tool | Shall have the capability to correctly calculate SUVs according to the vendor-neutral pseudo-codes for SUV calculation given in Appendix G. |
| Volume of Distribution Surrogate | Analysis Tool | Shall have the capability to calculate SUVs using as a surrogate for the Volume of Distribution: body weight, lean body mass, and body surface area (BSA). Lean body mass shall be calculated according to the formula of James [James 1976, Hallynck 1981]: Males: LBM = 1.10(w) − 128(w2/h2); Females: LBM = 1.07(w) − 148(w2/h2). Body surface area shall be calculated according to the Du Bois formula: BSA (m2) = (0.007184)((w)0.425)((h)0.725) [Vu 2002], where w = weight in kg and h = height in cm. |
4.4.3 Image Analysis Workstation Performance Specifications
The digital reference object (DRO), a synthetic PET (and CT) image, shall be used to evaluate conformance to the level of performance of the analysis/display station. Users should use the DRO (per the DRO user’s guide in Appendix F) to verify correct implementation of ROI placement, SUV calculations, and PET and CT image alignment.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Performance Evaluation | Analysis Workstation | Shall use the DRO to verify adequate performance as described in Appendix F. |
| Analysis Accuracy | Analysis Workstation | For each of the specified ROIs in the DRO (Appendix F) the correct SUV values shall be replicated by the Analysis Workstation. |
| Alignment Accuracy | Analysis Workstation | The PET and CT DRO object shall appear perfectly aligned in the transverse, coronal, and sagittal views. |
| DICOM Conformance | Analysis Workstation | Shall be able to read and apply all mandatory DICOM PET IOD attributes, as well as any additional optional DICOM attributes specified in this profile (including those private attributes defined in Annex G for SUV calculation). |
4.5. Software Version Tracking
Ideally, the PET/CT scanner should build a list on the console of the dates of all software versions (software changes that might impact quantitative accuracy would typically be inclusive of hardware change). The scanner software version should be identified and tracked across time, with updates and changes noted during the course of the trial. At a minimum, software versions should be manually recorded during qualification along with the phantom imaging performance data, and the record updated for every software upgrade over the duration of the trial.
| Parameter | Entity/Actor | Specification |
|---|---|---|
| Software Version tracking | Acquisition Device | Shall record the software version(s) used for acquisition and reconstruction in appropriate DICOM field(s). |
| Software version back-testing compatibility | Workstation | Shall provide mechanism to provide analysis of the image data using updated as well as prior (platform-specific) versions of analysis software. |
Appendices (Summary)
Appendix B: Background Information for Claim
A number of publications report test-retest repeatability for tumor SUV measurements with FDG-PET. Comparing repeatability across reports is complicated by differing methodologies and metrics. To allow direct comparison, the Committee translated reported repeatability measurements to a within-subject coefficient of variation (wCV). Based on the most directly comparable SUVmax studies (Nakamoto 2002, Krak 2005, Velasquez 2009, Hatt 2010, Weber 2015), the within-subject coefficient of variation for SUVmax was in the range of 10.0–12.0%, which underpins the Profile’s (−28%, +39%) repeatability claim.
Appendix D: Model-specific Instructions and Parameters
Appendix D provides vendor- and model-specific acquisition parameter tables and QA/QC procedure schedules (for representative GE, Philips, and Siemens PET/CT scanners) expected to produce data meeting the requirements of Section 3.6.4. The presence of a specific product model in these tables does not imply full compliance with the QIBA Profile; conformance involves meeting the full set of requirements. Sites using models not listed may devise their own acquisition parameters that result in conforming data. (These vendor grids are maintained in the source PDF and are omitted here for readability.)
Appendix F: FDG-PET/CT Digital Reference Object (DRO)
The PET/CT DRO is a synthetically generated set of DICOM image files of known voxel values for PET and CT, intended to test the computation of SUVs by PET/CT display stations and to test ROI calculations and PET–CT alignment. The CT object is 512 × 512 × 110 voxels; the PET object is 256 × 256 × 110 voxels; both have a reconstruction diameter of 500 mm and axial extent of 220 mm. In the PET object, voxels interior to the phantom body are set to an SUV of 1.00, the six hot spheres to an SUVbw of 4.00, and the shell/exterior/lung insert to 0.00. Two test voxels in slice 40 are set to SUVbw values of 4.11 and −0.11 (a value below zero being possible with analytic filtered back projection). Users import the PET and CT objects, perform ROI analyses on six specified ROIs, and submit results.
Appendix G: Vendor-neutral Pseudo-codes for SUV Calculation
Appendix G provides consensus generic and robust pseudo-codes for computing SUV from PET DICOM images. In outline: SUV is computed only if the Corrected Image attribute contains ATTN and DECAY and Decay Correction is START, and Units are BQML. The half-life is read from the Radionuclide Half Life; decay time is the difference between scan time and radiopharmaceutical start time; the decayed dose is the injected dose scaled by 2^(−decay time / half life); and SUVbw = (stored pixel value × Rescale Slope + Rescale Intercept) × (weight × 1000 / decayed dose), yielding g/ml. It is strongly recommended not to use Series Date and Series Time for decay correction. The most up-to-date version is maintained on the QIBA FDG-PET Wiki.
Appendix H: Consensus Formula for Lean-Body-Mass Normalization
Two formulas for estimating male LBM have been used in the PET community, differing in the coefficient (128 vs 120). The Profile recommends the James/Hallynck form – LBM(male) = 1.10 × Weight − 128 × (Weight/Height)2 and LBM(female) = 1.07 × Weight − 148 × (Weight/Height)2 (weight in kg, height in cm). The difference is minor at normal BMI but grows at high BMI. Alternative Janmahasatian equations have achieved some acceptance and may be recommended in future versions; consistency and standardization are expected to yield larger improvements in study power than potential improvements in LBM estimation accuracy.
Appendix I: QIBA FDG PET/CT Imaging Site Checklist
The following checklist may be used to ascertain a PET imaging site’s qualification for quantitative imaging according to the QIBA FDG PET/CT Profile. Answers may be provided either as “current practice” or as “feasible”, depending on the context, but it should be made clear both which was expected and how the site answered.
| # | Requirement | Conforms? |
|---|---|---|
| Site and Personnel Qualifications | ||
| 1 | The site is accredited (ACR, IAC, TJC, etc.) or has Qualified status for clinical trials (ECOG-ACRIN, SNMMI-CTN, EARL, CROs, etc.). | |
| 2 | The site has the support of technologists, physicists, and physicians experienced in the use of FDG-PET/CT, and meeting the qualifications described below. | |
| 3 | Technologists: PET studies are performed by technologists whose certification is equivalent to the recommendations published by representatives from the SNMTS or the ASRT; or certified as a nuclear medicine technologist in the country where the study is conducted; and should also meet all local, regional, and national regulatory requirements for the administration of ionizing radiation to patients. | |
| 4 | Physicists: The medical physicist is certified in Medical Nuclear Physics or Radiological Physics by the ABR; in Nuclear Medicine Physics by the ABSNM; in Nuclear Medicine Physics by the Canadian College of Physicists in Medicine; or certified in medical physics in the country where the study is conducted; or has 3 years of PET experience. Regardless of certification, the physicist should have specific experience in PET and its quantitative use. | |
| 5 | Physicians overseeing and interpreting PET/CT scans are qualified by the ABR (Diagnostic and/or Nuclear Radiology) or ABNM or certified as a nuclear medicine physician in the country where the study is conducted and/or interpreted. | |
| Imaging Procedures | ||
| 6 | Patient height and weight are measured and entered into the scanner during PET/CT acquisition. | |
| 7 | Blood glucose is measured for each patient within 2 hours preceding FDG administration. Measured value and measurement time are documented. | |
| 8 | Protocol-specific and institutional limits for the acceptable range of glucose are followed. If and when the glucose threshold is exceeded, the reason shall be documented. | |
| 9 | For each patient, the pre-injection FDG activity is measured, and injected and residual activity are measured. Initial and residual measurement times and injection time are entered into the console. If the scanner console is not capable of recording residual activity and measurement time, these should be documented separately. | |
| 10 | FDG is administered through a 24-gauge or larger indwelling catheter placed anatomically remote to any sites of suspected pathology, preferably in an antecubital vein. Intravenous ports should not be used unless no other venous access is available. In the case of manual administration, a three-way valve or alternative flush device should be attached to the intravenous cannula so as to allow at least a 10 cc normal (0.9% NaCl) saline flush following FDG injection. For automated injection devices, alternate flushing mechanisms are allowed. | |
| 11 | For follow-up scans, patients are imaged with the same workflow (i.e., patient handling, imaging acquisition, image processing, and image analysis) as for baseline scans. | |
| 12 | The FDG uptake time (from injection to scan) is 60 minutes, with an acceptable range of 55–75 minutes. When repeating a scan on the same subject who had a prior baseline scan outside the acceptable range, uptake time for the 2nd scan is within 10 minutes of that for the first scan. | |
| 13 | If the patient is observed to take a deep breath during the CT scan it is documented and a repeat CT study is considered. | |
| 14 | When a patient is rescanned, the same scan direction is used. | |
| 15 | Reconstructed PET images, with and without attenuation correction, and CT images are archived at the imaging site. | |
| QA/QC | ||
| 16 | The site performs all PET/CT scanner QA/QC procedures recommended by the manufacturer and at the recommended frequency (e.g., daily, weekly, quarterly) and assures that the output values are acceptable. | |
| 17 | Daily QA procedures are performed prior to any subject scan. | |
| 18 | A water or water-equivalent CT phantom is scanned and evaluated daily and acceptable output is ensured. Sites only performing CT for attenuation correction may perform this test weekly. | |
| 19 | Dose calibrator constancy is evaluated daily on the F-18 setting. Day-to-day differences from 2.5% to 5% should be investigated; differences no greater than 5% are allowed. Cs-137, Co-57, or simulated F-18 may be used. | |
| 20 | The dose calibrator accuracy is evaluated monthly with measured values differing no more than 5% from the actual source value. Values differing by 2.5% to 5% should be investigated. Cs-137, Co-57, or simulated F-18 may be used. | |
| 21 | Dose calibrator linearity is assessed at least quarterly over a range of 1–30 mCi (37–1110 MBq), with deviation of no more than 5% over the entire range. Values differing by 2.5% to 5% should be investigated. | |
| 22 | Scales for patient weight measurement are evaluated annually or after any repair by qualified personnel, with error no more than 2.5% from expected values using a NIST-traceable or equivalent standard. | |
| 23 | The glucose measuring device is measured and tested according to a CLIA-approved, CLIA-cleared, or equivalent (if outside the United States) procedure. | |
| 24 | The PET/CT scanner computer and all clocks in the imaging facility used to record activity/injection measurements are synchronized to standard time reference within ±1 minute. Synchronization is checked weekly and after power outages or civil changes for Daylight Saving (North America) or Summer Time (Europe). | |
| 25 | Quantitative Calibration Accuracy: PET scanner quantitative accuracy relative to the dose calibrator is verified quarterly and after scanner upgrades, maintenance or repairs, new setups and modifications to the dose calibrator via a uniform phantom scan of activity measured in the dose calibrator, achieving a large central ROI mean SUV value of 1.0 (acceptable range 0.95–1.05). | |
| 26 | Axial Uniformity: Using a uniform cylinder phantom or equivalent shall obtain a slice-to-slice variability of less than 10% for the slices within the central 80% of the axial FOV. | |
| 27 | PET Resolution: Cold rods (as in the Jaszczak or ACR PET phantoms) of diameter 9.5 mm or smaller must be visible. A hot cylinder (as in the ACR PET phantom) of 12 mm or smaller must be visible at 2.5:1 contrast OR the 10 mm sphere of the NEMA image quality phantom must be visible at 4:1 contrast. Alternative contrast ratios for specific accreditation requirements are allowed. | |
| 28 | PET noise: In a uniform phantom of 0.1 to 0.2 µCi/ml (3.7–7.4 kBq/ml) F-18 concentration, the coefficient of variation of voxel values within a rectangular or circular region of at least 3 cm (side or diameter) must be no greater than 15% for all slices within the central 80% of the axial FOV. | |
| Specific Personnel Responsibilities | ||
| 29 | A technologist or physicist assesses uniformity (within-plane and across slices) and compares with previous results. Quarterly and following software upgrades. | |
| 30 | A technologist or physicist shall perform the Quantitative Calibration Accuracy test. Quarterly and following software upgrades or changes to the dose calibrator. | |
| 31 | A physicist shall perform and document a quantitative assessment (using a phantom with differing size defined targets such as the ACR or NEMA IQ phantoms processed with routine image reconstruction protocols) for lesion resolution. Annually. | |
| 32 | A physicist shall perform a quantitative assessment of image noise in phantom images to be of consistent and acceptable quality. Annually. | |
Appendix I: QIBA FDG PET/CT Scanner Checklist
The following questionnaire/checklist may be used to ascertain a PET scanner’s qualification for quantitative imaging according to the QIBA FDG PET/CT Profile.
| # | Parameter | Specification | Conforms? |
|---|---|---|---|
| 1 | Calibration factors | All necessary calibration factors needed to output PET images in units of Bq/ml shall be automatically applied during the image reconstruction process. | |
| 2 | PET Scanner calibration | Shall be able to be calibrated according to the following specifications: using a uniform cylinder containing F-18 in water solution (ideally the same solution used for dose calibrator cross-calibration). Slice-to-slice variability shall be no more than ±5% (not including end slices, as per ACR PET Core Lab). | |
| 3 | Weight | Shall be able to record patient weight in lb or kg as supplied from the modality worklist or operator entry into scanner interface. Shall be stored in Patient Weight field (0010,1030) in the DICOM image header, as per DICOM standard. | |
| 4 | Height | Shall be able to record patient height in feet/inches or cm/m as supplied from the modality worklist or operator entry into scanner interface. Shall be stored in Patient Size field (0010,1020) in the DICOM image header, as per DICOM standard. | |
| 5 | Administered Radionuclide | Shall be able to enter the radionuclide type (i.e., F-18) by operator entry into the scanner interface and through predefined protocol. Shall be recorded in Radionuclide Code Sequence (0054,0300) in the DICOM image header [e.g., (C-111A1, SRT, “18Fluorine”)]. | |
| 6 | Administered Radiotracer | Shall be able to record the radiotracer (i.e., FDG), as supplied by operator entry into the scanner interface. Shall be recorded in Radionuclide Code Sequence field (0054,0300) in the DICOM image header, e.g., (C-B1031, SRT, “Fluorodeoxyglucose F18”). | |
| 7 | Administered Radiotracer radioactivity | Shall be able to enter the administered radioactivity, in both MBq and mCi, as supplied by operator entry into the scanner interface. Shall be recorded in Radionuclide Total Dose field (0018,1074) in the DICOM image header in Bq. | |
| 8 | Administered Radiotracer Time | Shall be able to record the time of the start of activity injection as supplied by operator entry into the scanner interface. Shall be recorded in Radiopharmaceutical Start Date Time field (0018,1078) (preferred) or Radiopharmaceutical Start Time field (0018,1072). | |
| 9 | Decay Correction Methodology | Encoded voxel values with Rescale Slope field (0028,1053) applied shall be decay-corrected by the scanner software (not the operator) to a single reference time (regardless of bed position), which is the start time of the first acquisition, encoded in the Series Time field (0008,0031) for original images. Corrected Image field (0028,0051) shall include the value “DECY” and Decay Correction field (0054,1102) shall be “START”. | |
| 10 | Scanning Workflow | Shall be able to support Profile Protocol (Section 3) PET and CT order(s) of acquisition. Shall be able to pre-define and save (by imaging site) a Profile acquisition Protocol for patient acquisition. | |
| 11 | CT Acquisition Parameters | Shall record all key acquisition parameters (technique) in the CT image header, using standard DICOM fields. | |
| 12 | PET-CT Alignment | Shall be able to align PET and CT images within ±2 mm in any direction. | |
| 13 | Activity Concentration in the Reconstructed Images | Shall be able to store and record (rescaled) image data in units of Bq/ml and use a value of BQML for Units field (0054,1001). | |
| 14 | Tracer Uptake Time | Shall be derivable from the difference between the Radiopharmaceutical Date Time field (0018,1078) (preferred) or Radiopharmaceutical Start Time field (0018,1072) and the Series Time field (0008,0031) or earliest Acquisition Time field (0008,0032) in the series, which should be reported as Series Time field (0008,0031). | |
| 15 | PET Voxel size | See Section 4.3 (PET Voxel size) under the Reconstruction Software specification requirements. | |
| 16 | CT Voxel size | Shall be no greater than the reconstructed PET voxel size. Voxels shall be square in transaxial dimensions, although are not required to be isotropic in the Z (head-foot) axis. Not required to be the same as the reconstructed PET voxel size. | |
| 17 | Subject Positioning | Shall be able to record the subject position in the Patient Orientation Code Sequence field (0054,0410) (whether prone or supine) and Patient Gantry Relationship Code Sequence field (0054,0414) (whether head or feet first). | |
| 18 | DICOM Conformance | All image data and scan parameters shall be transferable using appropriate DICOM fields according to the DICOM conformance statement for the PET/CT scanner. | |
| 19 | DICOM Data transfer and storage format | PET images shall be encoded in the DICOM PET or Enhanced PET Image Storage SOP Class, using activity-concentration units (Bq/ml) with additional parameters stored in public DICOM fields to enable calculation of SUVs. PET images shall be transferred and stored without any form of lossy compression. | |
| 20 | Metadata | Shall be able to accurately propagate the information collected at the prior stages and extend it with those items noted in the Reconstruction section. | |
| 21 | Data Corrections | PET emission data must be able to be corrected for geometrical response and detector efficiency, system dead-time, random coincidences, scatter and attenuation. | |
| 22 | Reconstruction Methodology | Shall be able to provide images without resolution recovery. | |
| 23 | Reconstruction Methodology / Output | Shall be able to perform reconstructions with and without attenuation correction. | |
| 24 | Data Reconstruction 2D/3D Compatibility | Shall be able to perform reconstruction of data acquired in 3D mode using fully 3D image reconstruction algorithms. Shall be able to perform reconstruction of data acquired in 2D mode using 2D image reconstruction algorithms. | |
| 25 | Quantitative calibration | Shall apply appropriate quantitative calibration factors such that all images have units of activity concentration, e.g., kBq/mL. | |
| 26 | Multi-bed data | Shall combine data from multiple over-lapping bed positions (including appropriate decay corrections) so as to produce a single three-dimensional image volume. | |
| 27 | Voxel size | Shall allow the user to define the image voxel size by adjusting the matrix dimensions and/or diameter of the reconstruction field-of-view. Shall be able to reconstruct PET voxels with a size 4 mm or less in all three dimensions (as recorded in Voxel Spacing field (0028,0030) and computed from the reconstruction interval between Image Position (Patient) (0020,0032) values of successive slices). Voxels shall be square in transaxial dimensions, although voxels are not required to be isotropic in the z (head-foot) axis. | |
| 28 | Reconstruction parameters | Shall allow the user to control image noise and spatial resolution by adjusting reconstruction parameters, e.g., number of iterations, post-reconstruction filters. | |
| 29 | Reconstruction protocols | Shall allow a set of reconstruction parameters to be saved and automatically applied (without manual intervention) to future studies as needed. |